TIERA Book demo ->
Technology blog
Envelope analysisAccelerometer selectionBearing faultsIMI SensorsInteractive simulator
Interactive · Sensor selection / 18 min + four simulators

Your Envelope Band Is Perfect. Your Sensor Can't Reach It.

A boiler feed pump destroyed a bearing three weeks after a route that read perfectly clean. Nothing was wrong with the analysis. Everything was wrong with the instrument. Four simulators take the failure apart — reach, noise floor, headroom and the loop — across the eight IMI precision ICP accelerometers.

01

Act I · A bearing that failed three weeks after a clean route

The machine is a boiler feed pump: 200 kW, 1,440 rpm, 24 Hz at the shaft, with a 6316 deep-groove ball bearing at the non-drive end. Outer-race defect frequency, 88 Hz. It sat on a monthly vibration route for four years without incident.

On the March route it read clean. Overall velocity 2.1 mm/s RMS, comfortably in ISO 20816 Zone A. The spectrum showed 1× shaft, a little 2×, vane pass, nothing else. The analyst signed it off, correctly, on the evidence in front of him.

Twenty-three days later the bearing seized. The strip-down found an outer-race spall roughly 9 mm across with established secondary damage around it — a defect that does not appear in three weeks. It had been growing for the better part of a year, through eight monthly routes, every one of which read clean.

The obvious conclusion is that somebody missed it. The obvious conclusion is wrong. Nobody missed it, because it was never in the data. The envelope band was tuned correctly, the analysis was competent, and the defect was physically incapable of reaching the analyser — for three separate reasons, each of which is printed on a datasheet nobody had read.

What the bearing was doing Defect energy in the resonance band, growing month on month DETECTION FLOOR OF THE INSTRUMENT ACTUALLY FITTED eight monthly routes — every one reported CLEAN SEIZED the defect crosses the floor only three weeks before failure The analysis was never the problem. The measurement floor was above the fault for eleven of the twelve months.
The failure mode of a condition-monitoring programme is rarely a missed diagnosis. It is a detection floor sitting above the fault for most of the fault's life — and a detection floor is a purchasing decision, not an analysis decision.
02

Act I · Three gates stood between the spall and the spectrum

Two earlier posts on this blog take the technique apart. Why a Failing Bearing Hides in the Noise explains the physics: a spall is a tiny, perfectly repetitive impact that rings the housing at a few kilohertz, and enveloping recovers its rhythm by band-passing that ringing and rectifying what is left. Tune the Band, Find the Bearing makes it a tuning exercise you can drive: park the window on the resonance and the comb appears; slide off and it vanishes.

Both posts quietly assume something that was false on this pump. They assume the signal reaches the analyser.

An envelope band is a request. The accelerometer is what answers it. If the housing rings at 6 kHz and the sensor is specified flat to 10 kHz but resonates at 18 kHz, you are working where the sensor is already inflating your amplitudes. If the spall puts 300 µg of ringing into the housing and the sensor's own electrical noise contributes 330 µg across the same band, no tuning on earth finds the comb — you would be demodulating the amplifier. And if the machine hammers hard enough to drive a ±50 g sensor into its stops, the clipping manufactures a comb of its own, at frequencies that have nothing to do with the bearing.

Reach, noise floor, headroom. Three gates, all decided before anyone opens the analyser, all printed on the datasheet. Act II takes them one at a time against a real product line — the IMI Sensors precision ICP range, eight series from the 621 to the 628 — and gives you an instrument for each.

What has to survive, in order Spall impact GATE 1 Reach Is the ringing band inside the sensor's specified response? Set by the resonance. GATE 2 Noise floor Is the impact bigger than the sensor's own noise in that band? Density × √bandwidth. GATE 3 Headroom Do the transients stay inside the measurement range, unclipped? Set by full scale. Shut any one of the three and the envelope spectrum is empty — or worse, convincingly wrong. None of them is a software setting.
Enveloping is usually taught as a filter-tuning skill. The filter is the last step. Three hardware gates come first, and all three are decided at the purchase order.
03

Act II · Gate one — could the sensor reach the ring?

A rolling-element defect does not radiate at its defect frequency. It radiates a short impulse, and that impulse excites whatever structural resonance is nearest — the bearing outer ring, the housing, the sensor's own mounted resonance. In industrial machines that ringing typically lands between 2 kHz and 20 kHz. Enveloping works precisely because that band sits far above the shaft and blade energy that swamps the low end.

So the sensor has to be honest up there. A piezoelectric accelerometer is a mass on a spring: below its mounted resonance it is flat, approaching resonance it lifts, past resonance it collapses. That is why the datasheet prints two different numbers, and confusing them is the most common selection error in the field.

Mounted resonance is where the sensor peaks. The ±3 dB corner is where the manufacturer will still guarantee the amplitude. They are not the same, and the gap is large: the 623 Series resonates at 40 kHz but is specified only to 15 kHz. Between the two it still produces output — often a great deal of it — but that output is no longer calibrated. It will find a defect. It will not tell you how big the defect is.

Plot the range against the band enveloping actually uses and the ranking falls out. The three quartz units — 624, 627, 628 — all resonate at 18 kHz and are specified to 10–12 kHz, so their rise begins exactly where the envelope band lives. The 623 (40 kHz) and 621 (85 kHz) stay genuinely flat across the whole of it.

ENVELOPING BAND — 2 to 20 kHz 0 dB +10 −10 500 1k 2k 5k 10k 20k 50k 100k Frequency (Hz, log) 621 — 85 kHz 623 — 40 kHz 622 — 30 kHz 624/627/628 — 18 kHz Dotted verticals: each model's published ±3 dB limit.
Single-degree-of-freedom response computed from each model's published mounted resonance (damping ratio 0.03). Peaks are clipped at +20 dB. The dotted verticals are the separate, more conservative ±3 dB specification limits — 10 kHz for the quartz units, 15 kHz for the 622 and 623, 30 kHz for the 621. Note what the grey dashed curve does inside the orange band: the quartz units are climbing exactly where the bearing rings.
04

Act II · Gate two — the arithmetic nobody does

This is the gate that killed the pump, and it is the one almost never checked, because it needs one line of arithmetic the datasheet does not do for you.

Every accelerometer's built-in amplifier contributes noise. The datasheet gives it as a spectral density — µg per root hertz — at 10 Hz, 100 Hz and 1 kHz. A density is not a level. To find out how much noise you actually get, integrate it across the bandwidth you are using. For a flat density that is simply:

noise in band = density × √bandwidth

An envelope band is wide by design — you want the whole resonance, so 2,000 to 4,000 Hz is typical. Take 3,000 Hz and run the eight models through it, using each one's density at 1 kHz, the right figure for a band sitting in the kilohertz region.

The spread is brutal. The 622 delivers 22 µg of noise across that band. The three quartz units deliver 329 µg — fifteen times more, 23.5 dB. An incipient outer-race spall putting a couple of hundred µg of ringing into the housing is a clean, obvious comb on the 622 and is literally below the noise on the 627. Same machine, same analyser, same perfectly tuned band, same analyst, opposite conclusions.

The pump had a 627 fitted. It had been specified years earlier, correctly, for its long-term stability in a hot pump house — and then quietly asked to do a job it was never built for.

There is a second surprise in the table, and it is the honest thing to say about the sensor with the best bandwidth. The 621 has the worst noise floor of the eight. Its 10 mV/g sensitivity buys the ±500 g range and the 85 kHz resonance, and pays for them with 10 µg/√Hz — 548 µg in our band, worse even than the quartz units. The 621 is not the answer to "find me a faint bearing defect." It answers a different question, which is Gate 3.

Sensor self-noise across a 3,000 Hz envelope band — density at 1 kHz × √3000 10 100 1000 Noise in band (µg RMS, log scale) 622B01 22 µg 625B01 27 µg 626B01 38 µg 623C01 49 µg 624 / 627 / 628 329 µg 621C40 548 µg The pump's 300 µg spall clear on the four low-noise ceramics ← below the floor on quartz and on the 621
The same defect, the same band, eight sensors. Derived from published spectral noise density at 1 kHz; the 622 and the 627 differ by a factor of fifteen. This is not a subtle difference in sensitivity — it is the difference between a diagnosis and a blank spectrum.
05

Act II · Gate three — and the comb that was never there

Gates 1 and 2 push you toward a sensitive, low-noise, high-resonance sensor. Gate 3 pushes back, and on some machines it wins outright.

Six of the eight models are ±50 g full scale. That is generous for a motor or a pump and not remotely generous for a hammer mill, a reciprocating compressor, a crusher, or a gearbox under shock load. Drive a ±50 g sensor past its range and it does not saturate quietly. It clips — and a clipped waveform is rich in harmonics that were never in the machine.

The simulator below computes this rather than asserting it. It synthesises a machine with a large 1× shaft vibration and an optional bearing defect, hard-clips the waveform at the selected sensor's full scale, then runs a real band-pass, a real Hilbert envelope and a real 4,096-point FFT of that envelope — the same chain a analyser runs.

Set the defect to zero and raise the 1× level past ±50 g. A comb appears. It is not at BPFO; it stands at twice shaft rate and its harmonics, because hard clipping puts a sharp corner into the waveform twice per revolution and the envelope detector reads those corners as impacts. There is no bearing fault anywhere in the signal. Switch the sensor to the ±500 g 621 and the comb vanishes — proving it was the instrument, not the machine.

Then set a real defect and clip it. Two things happen at once, and the second is the one that costs money: a false comb appears at 2× shaft, and the real BPFO comb is suppressed, because flat-topping compresses exactly the impact peaks the envelope is measuring. Clipping invents a fault and understates the real one in the same record.

Interactive — drag the controls

The pump: 1,440 RPM (24 Hz shaft), 6316 bearing, BPFO 88 Hz, housing ringing near 3 kHz. Record: 4,096 samples at 8,192 S/s. Band-pass 2–4 kHz, Hilbert envelope, 4,096-point FFT — all computed live in your browser.

Start on the default — no defect at all, 80 g of 1× on a ±50 g sensor — and read the blue comb. It is entirely manufactured by the clipping. Press "Real defect, no clipping" to see what a genuine BPFO comb looks like in orange, then "Both at once" to watch the false comb rise while the real one is squashed. Switching to the ±500 g 621 removes the artefact without touching the machine.
06

Act II · The instrument that settles all three gates

Each gate is one calculation, so here they are as a single instrument. Pick a sensor, describe the machine and the mount, and watch all three verdicts move together.

The response curve is computed from each model's published mounted resonance as a single-degree-of-freedom system, with the mount cascaded on top. The noise line is that model's published 1 kHz density integrated across the band width you choose. The envelope spectrum at the bottom is drawn relative to that noise floor — so when the comb sinks into the grass, it is sinking into this specific sensor's grass, for the reason named in the panel.

Three things worth trying. Leave the ring frequency at 3.5 kHz on a stud mount and step through all eight models: the response curve barely moves, and the SNR readout alone swings from +22.7 dB on the 622 to +15.7 dB on the 623, −0.8 dB on the 627 and −5.2 dB on the 621. Same machine, same band, same fixture — Gate 2 decides the entire outcome by itself. That single readout is the pump's whole story.

Then leave the 622 selected and switch the mount from stud to magnet, and watch something genuinely counter-intuitive. The SNR gets better — from +22.7 dB to +32.1 dB — because at 3.5 kHz you sit just below the magnet's own 4.2 kHz resonance and it is handing you 9.5 dB of gain. Gate 1 turns amber anyway, and it is right to: that gain is not calibrated, it moves with every remount, and the amplitude you trend has become a property of the magnet. Now drag the ring frequency upward and watch the same fixture turn on you — the mount contributes −8.5 dB at 8 kHz, −17 dB at 12 kHz and −27 dB at 20 kHz, and Gate 1 goes red the moment you pass its 4.2 kHz resonance.

Finally set the transient to 200 g and watch seven of the eight models fail Gate 3 at once, leaving the 621 — the model with the worst noise floor in the range — as the only one that can take the measurement at all.

Interactive — drag the controls

Bearing: outer-race defect, BPFO = 88 Hz — the pump from Act I, and the same case as the tuning simulator. All eight models carry their published brochure specifications.

GATE 1 · REACH

GATE 2 · NOISE FLOOR

GATE 3 · HEADROOM

Response is modelled as a single-degree-of-freedom system from each model's published mounted resonance (sensor damping 0.03), with the mount cascaded on top — so the orange curve is sensor × fixture, which is what you actually measure with. The noise line is that model's published 1 kHz density integrated over the band width you set. Watch Gate 1 and Gate 2 disagree when you fit a magnet: better SNR, worse measurement.
07

Act III · How much warning does each sensor buy?

Decibels are an abstraction. A maintenance manager does not buy decibels; they buy time — the interval between the first credible warning and the outage. That interval is what the noise floor actually sets, and it is straightforward to compute.

Bearing defect energy grows roughly exponentially once a spall initiates. Give a degradation curve a starting level and a failure level, and each sensor's detection threshold — its own in-band noise, times whatever confidence margin you require — intersects that curve at a different date. The gap between that date and failure is the warning time you have bought.

The instrument below does exactly that. It is the same arithmetic as Gate 2, replotted against a calendar instead of a spectrum.

On the default settings the answer is stark. The 622 raises a credible flag with something over a year in hand. The 627 that was actually fitted to the pump gets there months later — and on a machine whose degradation runs faster than this one, "months later" becomes "three weeks before it seizes," which is precisely what happened.

Two honest caveats sit on this chart, and both are stated on it. Real degradation is not a clean exponential — bearings plateau, and a self-honing spall can look like recovery for weeks. And detection is not the same as diagnosis: crossing the noise floor makes the comb visible, not yet trendable. Both push the real warning time down from what the model says, which is the safe direction for a chart used to justify a sensor.

Interactive — drag the controls

Defect energy in the resonance band grows exponentially from onset to failure. Each sensor is flagged when the growing signal first exceeds its own in-band noise by the confidence margin you set.

SENSORIN-BAND NOISEFIRST DETECTEDWARNING
The vertical axis is logarithmic, which is why exponential growth plots as a straight line. Each dashed line is one sensor's detection threshold; the dot is the month it first clears. On the defaults the 622 flags at month 5.7 with a year in hand, while the quartz units wait until month 11.6 — that gap is what 0.4 µg/√Hz buys over 6 µg/√Hz. Selecting a magnet appears to add warning time for every sensor at once; that is the mount's uncalibrated 9.5 dB of gain, not a better measurement, and Gate 1 in the previous simulator rejects it.
08

Act III · The loop between the sensor and the analyser

There is a fourth failure that belongs with the other three, because it also caps your bandwidth and it also never appears in an analysis review. It is the two-wire loop itself.

Every model in this range is an ICP® sensor: a constant-current, two-wire system with a built-in amplifier. The brochure sets out the operating conditions precisely. An 18–30 VDC supply feeds a current-regulating diode that holds the line between 2 and 20 mA, dropping about 1 V across itself. The sensor's amplifier sits on that current and produces a DC bias voltage, normally 8 to 12 VDC, with the AC vibration signal riding on top. A 10–30 µF decoupling capacitor at the conditioner strips the DC away so the analyser sees only the signal.

That bias voltage is a free diagnostic and almost nobody reads it. A healthy sensor sits mid-rail. An open circuit — a broken conductor, a connector backed out, a sensor not fitted — lets the bias float up to the supply rail. A short to shield pulls it to zero. Moisture in a connector drags it down and makes it wander. You can tell all of this from a single DC measurement before you record anything.

The second thing the loop decides is more surprising, and it matters directly to everything above: the constant current and the cable capacitance together set a hard ceiling on frequency, and that ceiling can land below the bearing's ring. Driving cable capacitance at high frequency and high amplitude requires current, and an ICP sensor has only what the conditioner supplies minus roughly 1 mA for its own bias:

fmax = 109 × (Ic − 1) / (2π C V) — with Ic in mA, C the total cable capacitance in pF, and V the peak signal in volts.

Work it through. A 4 mA conditioner, 100 m of cable at 100 pF/m and a 5 V peak signal gives 9.5 kHz — enough for a 3.5 kHz housing resonance, not enough for the top of the band. Drop the conditioner to the 2 mA minimum and the same run gives 3.2 kHz, which lands directly on the resonance you were trying to measure. Take a 200 m run of 150 pF/m cable at 2 mA with an 8 V signal and you are down to 660 Hz: the enveloping band is simply absent, and the spectrum will look perfectly plausible while it is missing.

It works the other way too. Raise the conditioner to 10 mA on that 100 m run and you reach 28.6 kHz; shorten the run to 10 m at 4 mA and you are at 95 kHz, far beyond any sensor in the range. This is exactly why the brochure specifies 2 to 20 mA rather than a single value — and why a long cable run is a bandwidth decision rather than a convenience one.

Interactive — drag the controls

ICP® two-wire loop, per the brochure's operating conditions: 18–30 VDC supply, 2–20 mA constant current through a regulating diode (about 1 V drop), sensor bias normally 8–12 VDC.

BIAS VOLTAGE

CABLE FREQUENCY CEILING

Two checks that cost nothing. The bias voltage clears the loop before you record — mid-rail is healthy, at the supply rail is an open circuit, on the floor is a short, and anything below 8 V is a connector you should open. The frequency ceiling is the one that surprises people: the default 100 m at 4 mA reaches 9.5 kHz, but pull the current back to the 2 mA minimum and it collapses to 3.2 kHz — right onto the housing resonance. Lengthen the run or raise the signal voltage and it drops through the floor.
09

Act III · Which one, for which job

Run the gates across the range and the eight series sort into four honest answers rather than a ranking. There is no best accelerometer here, only a best fit — and the quartz units are not the losers of this comparison. They are the answer to a question the pump was not asking.

The default for bearing enveloping: 623. A 40 kHz mounted resonance keeps it flat through the whole practical ringing band, 0.9 µg/√Hz gives 49 µg across a 3 kHz window, and at 51 g and 17.3 mm diameter it fits pump and motor housings where bigger ring-style bodies will not go. It clears Gates 1 and 2 together without an argument.

When the defect is faint: 622. The lowest noise floor in the brochure — 0.4 µg/√Hz at 1 kHz, 50 µg broadband — buys the earliest possible detection, which is the entire point of a trending programme and precisely what the warning-time chart above prices. You trade the 623's resonance headroom (30 kHz rather than 40 kHz) for roughly 7 dB of visibility. On a machine ringing at or below 5 kHz that is a straight win. The 625 is the same argument in a side-exit ring-style body when cable routing is tight, at 0.5 µg/√Hz.

When the machine hammers: 621. ±500 g, a 10,000 g shock limit, ±3 dB to 30 kHz and 4.75 grams. Gate 3 overrules Gates 1 and 2 on crushers, mills, reciprocating machines and shock-loaded gearboxes, because a clipped record is not a worse measurement — the clipping simulator above shows it is a different and wrong one. It is also the only model here that reaches gear-mesh harmonics well past 20 kHz.

When the environment is the problem: 624, 627, 628. Quartz is naturally piezoelectric and does not depolarise, so it holds sensitivity over years and across temperature swings in a way ceramic does not. That is a real and valuable property — for permanent installations in thermally active plant, for long trend records that must stay comparable, for paper mills, conveyors and outdoor duty. It is simply not the property that finds a faint spall: 1,000 µg broadband resolution and 6 µg/√Hz put these units 15× above the 622 in noise. Specify them for stability, and do not ask them to do incipient enveloping. That single mismatch is the whole of Act I.

The special case: 626. A larger seismic mass for slow-turning machinery, buildings and structures, with 15 µg/√Hz at 1 Hz. Enveloping is rarely the technique down there, but when a slow-speed bearing does need it, the 626 still has signal at the shaft rate to reference the comb against.

All figures from the IMI Sensors precision ICP brochure. Noise-in-band is derived: density at 1 kHz × √3000. MODEL mV/g RANGE ±3 dB RESONANCE µg/√Hz IN 3 kHz ENVELOPING VERDICT 623C01 100 ±50 g 15 kHz 40 kHz 0.9 49 µg Default choice 622B01 100 ±50 g 15 kHz 30 kHz 0.4 22 µg Earliest detection 625B01 100 ±50 g 10.5 kHz 25 kHz 0.5 27 µg Side exit, tight cable runs 621C40 10 ±500 g 30 kHz 85 kHz 10 548 µg Shock & gear mesh only 626B01 100 ±50 g 10 kHz 23 kHz 0.7 38 µg Slow-speed machines 624B01 100 ±50 g 10 kHz 18 kHz 6 329 µg Stability duty, not enveloping 627A01 100 ±50 g 10 kHz 18 kHz 6 329 µg Stability duty — the pump's mistake 628F01 100 ±50 g 12 kHz 18 kHz 6 329 µg Outdoor / mill duty Every model shares: shear-mode sensing, welded hermetic stainless housing, IP68, −54 to +121 °C, 18–28 VDC / 2–20 mA ICP, ±1 % non-linearity and ≤5 % transverse sensitivity. EX, HT, metric-mount, temperature-output and velocity-output variants exist.
The verdict column is the point. Four of the eight are genuinely good enveloping sensors, three are excellent sensors for a job that is not enveloping, and one is the shock specialist. A range is not a ladder.
010

Act III · The gate that is on nobody's datasheet

You can clear every gate on paper and still get nothing, because one constraint no accelerometer specification can protect you from is the interface between the sensor and the machine.

The mounted resonance printed on a datasheet is measured on a stud, into a prepared, flat, clean spot face, torqued to specification. Change the fixture and you change the number. Our mounting bandwidth simulator puts practical ceilings at roughly 10 kHz for a stud, 5 kHz for a thin hard adhesive bond, 1.4 kHz for a flat magnet on clean bare metal and 500 Hz for a handheld probe. Read those against the 2–20 kHz band this whole post is about.

It is worth being precise about how a magnet fails, because the naive version of the warning is wrong and the gate simulator will show you so. A flat magnet resonates around 4.2 kHz. Below that it does not attenuate at all — it amplifies, by about 9.5 dB at 3.5 kHz. Your comb gets taller and your trend gets louder, and none of it is calibrated: the gain depends on the surface, the film of dirt and exactly where the magnet last landed, so the same bearing measured twice gives two different severities. Above its resonance the magnet collapses properly — roughly −8.5 dB at 8 kHz, −17 dB at 12 kHz, −27 dB at 20 kHz — and there the band really is gone.

So a magnet gives you one of two failures depending on where the housing rings: an inflated number you cannot trend, or no signal at all. Neither is a measurement. The pump had both problems at once — a quartz sensor on a two-pole magnet — which is why eleven months of routes produced eleven months of nothing. The argument is made at length in why cheap cables and magnets cost more.

The brochure is specific about the fixture and it is worth honouring. The 621 uses a 5-44 male thread at 2.0 to 2.2 N·m. The 622, 623, 624, 625, 626, 627 and 628 all use ¼-28 at 2.7 to 6.8 N·m, with an M6×1 metric-mount variant available across the range. Every one is welded hermetic, IP68 and rated −54 to +121 °C, with an HT prefix taking the 622, 623, 624, 625 and 628 to +163 °C and an EX prefix certifying the 621, 622, 623, 625 and 628 for ATEX, CSA and IECEx hazardous areas.

The last link is the cable that carries a signal measured in tens of µg through an electrically filthy plant. A low-noise coaxial cable is not an accessory for that job; a triboelectric-noisy one contributes more than the 622's entire noise budget. That argument is made in how vibration cables are tested, and the recording end of it in what makes a DAQ trustworthy.

where bearings ring — 2 to 20 kHz Stud 10 kHz Adhesive pad 5 kHz Flat magnet 1.4 kHz Handheld probe 500 Hz 500 2k 5k 10k 20k Two of the four fixtures never reach the band at all.
Log frequency axis; bars mark the usable ceiling, above which the fixture — not the sensor — sets what you can believe. Two of the four never reach the enveloping band. The best accelerometer in this post, on a magnet, is a worse enveloping instrument than the cheapest one on a stud.
011

Act IV · What the plant does now

The pump was rebuilt with a 623C01 stud-mounted into a spot face machined onto the bearing housing, on 12 m of low-noise coaxial cable into a 10 mA conditioner. Bias voltage is recorded at the start of every route as a single line in the sheet. Nothing about the analysis changed; the analyst was never the problem.

The habit worth breaking is buying an accelerometer on sensitivity. 100 mV/g tells you almost nothing about whether a bearing fault will be visible, because sensitivity is a scaling factor and detection is a ratio. Six of the eight models here are 100 mV/g and they differ by fifteen times in what they can actually resolve.

Read the datasheet in this order instead. Mounted resonance — does the sensor reach the ring? Spectral noise density — can it see the ring once it gets there? Measurement range — will the record survive the machine? Three numbers, in that order, and an eight-model brochure collapses into a shortlist in about a minute.

Then check the fixture, because none of it survives a magnet. And check the loop, because none of it survives 100 m of cable at 4 mA.

The uncomfortable part of this case is that nothing in it was exotic. A correct sensor, specified for the wrong job, on a convenient mount, at the end of a long cable. Every one of those decisions was made by someone reasonable, and none of them was made by the person who later got asked why the bearing failed.

The kit for this job

TIERA instruments that do this work.

IMI 623C01 — High Frequency ICP® Accelerometer

IMI 623C01 — High Frequency ICP® Accelerometer

The default bearing-enveloping sensor, and what the pump runs now: a 40 kHz mounted resonance keeps it flat across the whole ringing band, and 0.9 µg/√Hz keeps the comb above its own noise. Small enough at 17.3 mm diameter and 51 g for crowded pump and motor housings.

Sensitivity
100 mV/g (±5 %)
Frequency range (±3 dB)
0.8 to 15,000 Hz
Mounted resonance
40 kHz
Spectral noise at 1 kHz
0.9 µg/√Hz
Broadband resolution
100 µg (1–10 kHz)
Mount / weight
¼-28 female, 2.7–6.8 N·m / 51 g
IMI 622B01 — Top Exit ICP® Accelerometer

IMI 622B01 — Top Exit ICP® Accelerometer

The lowest noise floor in the precision range — 22 µg across a 3 kHz envelope band, against 329 µg for the quartz units. When the value of the programme is in catching the defect early, this is the sensor that buys the months.

Sensitivity
100 mV/g (±5 %)
Frequency range (±3 dB)
0.2 to 15,000 Hz
Mounted resonance
30 kHz
Spectral noise at 1 kHz
0.4 µg/√Hz
Broadband resolution
50 µg (1–10 kHz)
Mount / weight
¼-28 female, 2.7–6.8 N·m / 94 g
IMI 621C40 — Very High Frequency ICP® Accelerometer

IMI 621C40 — Very High Frequency ICP® Accelerometer

For machines that hammer. ±500 g full scale and a 10,000 g shock limit keep the record unclipped where a ±50 g sensor manufactures the false comb the simulator above computes — and ±3 dB to 30 kHz reaches gear-mesh harmonics nothing else here can see, in 4.75 grams.

Sensitivity
10 mV/g (±10 %)
Measurement range
±500 g
Frequency range (±3 dB)
1.6 to 30,000 Hz
Mounted resonance
85 kHz
Overload limit (shock)
10,000 g pk
Mount / weight
5-44 male, 2.0–2.2 N·m / 4.75 g
PhonoVibe Series — Sound & Vibration DAQ

PhonoVibe Series — Sound & Vibration DAQ

A sensor that reaches 15 kHz is wasted on a recorder that does not. Q/O/HD sample at 128 kHz across 0.5 Hz – 60 kHz with 24-bit resolution and built-in ICP/IEPE power, so the ringing band survives the whole chain.

ADC resolution
24-bit, simultaneous sampling
Sampling (Q/O/HD)
128 kHz
Bandwidth (Q/O/HD)
0.5 Hz – 60 kHz
Sensor power
24 V, 4 mA (IEPE/ICP/CCLD)

From the TIERA store

The kit for this job

What we would actually put in front of someone doing the measurement this post describes — not the whole catalogue.

Use cases

Where this shows up in the field

From TIERA

Tell us the machine — we will tell you which of the eight, and why

TIERA specifies and supplies the IMI Sensors precision ICP range alongside our own PhonoVibe DAQ, TVIB analysis software, and the mounting and cabling that decide whether any of it works. The gates in this post are exactly how we make the recommendation: what the housing rings at, how faint the defect will be when you first want to catch it, how hard the machine hits, and what the cable run and conditioner will actually carry. Give us the machine list — speeds, bearing part numbers, duty, cable runs — and the shortlist usually comes down to one or two models per measurement point.

The models are also available with the prefixes the plant may need: EX for ATEX, CSA and IECEx hazardous areas on the 621, 622, 623, 625 and 628; HT to raise the operating range from +121 °C to +163 °C on the 622, 623, 624, 625 and 628; M for M6×1 metric mounting across the whole range; TO for a built-in temperature output; and VO for a velocity output at 100 mV/ips. Alternate sensitivities of 10, 50 and 500 mV/g are available on most models, and every series offers integral polyurethane or armoured-polyurethane cable versions with configurable length and terminating connector.

And if you want the pattern recognition before the plant provides you with real failures, the TMFSS simulator seeds outer-race, inner-race, cage and rolling-element defects under controlled speed — so the comb in these simulators is one you have already measured with your own hands, through your own sensor, at a severity you chose.

  • IMI precision ICP accelerometers, 621 to 628 — specified against your machines, with EX / HT / metric / temperature-output / velocity-output variants
  • PhonoVibe Q / O / HD: 24-bit, 128 kHz sampling, 0.5 Hz – 60 kHz, ICP sensor power on every channel — the recording end of the same bandwidth argument
  • TVIB TSAP201 (free with every PhonoVibe): FIR/IIR band-pass, FFT to 102,400 points, harmonic and sideband cursors for laying a ruler over the comb
  • Stud-mount blocks, adhesive pads and low-noise coaxial cable — the two gates no datasheet protects you from
  • TMFSS benchtop fault simulator with seeded bearing defects, for practising the whole chain before it matters
Learn this properly

Where this sits on the TIERA learning ladder.

The theory behind this article is covered free, in full, by the TIERA 101 primers: Accelerometer & DAQ Selection 101, Bearing & Gear Analysis 101. They are self-paced, interactive, and end in an exam and a certificate.

TCAT Cat I-IV adds structured analyst training with proctored examinations at exams.tieraonline.in — the free 101 primers cover the theory, TCAT certifies you can apply it.

TIERA 101 is a free introductory primer, not an accredited ISO certification, and its hours do not count towards the formal training ISO 18436 requires.